Introduction:
The ultrasonic flow meter is designed to measure the fluid velocity of liquid within a closed conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of non-fouling operation and easy installation.
The flow meter operates by alternately transmitting and receiving a frequency modulated burst of sound energy between the two transducers and measuring the transit time that it takes for sound to travel between the two transducers.
Parameter:
|
Performance &Parameter |
Convertor
|
Principle |
Transit-time ultrasonic flowmeter |
Accuracy |
±1% |
Display |
2x20 character LCD with backlight, support the language of Chinese, English, Italy, Turkish |
Signal output |
1way 4-20mA output, electric resistance 0-1K, accuracy0.1% |
1 way OCT pulse output(Pulse width-1000ms, default is 200ms)) |
1 way Relay output |
Signal input |
3 way 4-20mA input, accuracy 0.1%, acquisition signal such as temperature, press and liquid level |
Connect the temperature transducer P100, can finish the heat/energy measurement. |
Data Interface |
Insulate Rs485 serial interface, upgrade the flowmeter software by computer, support the MODBUS. |
Special Cable |
Twisted-pair cable, generally, the length under 50 meters, select the Rs485, transmission distance can over 1000m |
Pipe Installation Condition |
Pipe material |
Steel, stainless steel, cast iron, copper, cement pipe, PVC, Aluminum, glass steel product, liner is allowed |
Pipe diameter |
25-100mm |
Straight pipe |
Transducer installation should be satisfied: upstream10D, downstream5D, 30D from the pump. |
Measuring Medium |
Type of liquid |
Single liquid can transmit sound wave; such as water (hot water ,chilled water, city water, sea water, waste water, etc); sewage with small particle con tent, oil(curde oil, lubricating oil, diesel oil, fuel oil, etc); chemicals(alcohol, etc); plant effluent; beverage; Ultra-pure liquids, etc |
Temperature |
-30-160¡É |
Turbidity |
No more than 10000ppm and less bubble |
Flow rate |
0-±7m/s |
Temperature |
Convertor; -20-60¡É; Flow Transducer: -30-90¡É/160¡É(Optional) |
Humidity |
Convertor: 855RH; Flow Transducer: can measure under water, water depth¡Â2m |
Working Environment |
Temperature |
Convertor; -20-60¡É; Flow Transducer: -30-90¡É/160¡É(Optional) |
Humidity |
Convertor: 855RH; Flow Transducer: can measure under water, water depth¡Â2m |
Power Supply |
DC8-36V or AC85-264V |
Power Consumption |
1.5W |
Dimension |
132*150*85mm(convertor) |
Terminals:
The following parameters need to be configured for a proper measurement:
(1) Pipe outer diameter
(2) Pipe wall thickness
(3) Pipe materials (for non-standard pipe materials*, the sound speed for the material must be configured too)
*Standard pipe materials and standard liquids refer to those with the sound parameters that have already been programmed into software of the flow meter, therefore there is no need to configure them
(4) Liner material and its sound speed and thickness, if there is any liner.
(5) Liquid type (for non-standard liquids, the sound speed of the liquid is also needed)
(6) Transducer type adapted to the flow meter.
(7) Transducer mounting methods (the V-method or Z-method is the common option)
(8) Check up the Space displayed on M25 and install the transducers accordingly.
(9) Store the parameter setup.
Requirements for pipeline installation:
Try to meet the requirements of the straight pipe section, at least 10 times the pipe diameter from the upstream elbow and 5 times the downstream pipe diameter. When not satisfied, install the sensor as close as possible to the lower part of the middle.
If there is an insulation layer outside the pipe, remove part of the insulation layer according to the installation distance and installation mode, and ensure that the two sensors are on the same axis section after positioning, and the sensor must be installed on the side of the pipe.
Metal pipe needs to be polished, use angle grinder or straight plate file in the installation point to grind out a square of about 1.5 times the area of the bottom of the sensor (PVC tube does not need grinding, wipe clean) after polishing requirements: bright, non-corrugated, no rust point.
Sensor installation and adjustment:
(1) At the bottom of the upstream (UP) sensor, evenly smear 3~5mm with a coupling agent or silica gel, and secure the upstream (UP) sensor at the pipe grinding point.
(2) View the sensor installation distance in the menu M25 window.
(3) At the bottom of the downstream (DN) sensor, evenly smear 3~5mm with a coupling agent or silica gel, and then slightly secure the downstream (DN) sensor at the distance shown by M25.
(4) Slowly move the downstream (DN) sensor and observe the M91 and M90 windows at the same time to ensure that the transmission ratio of the M91 window is less than 100 ±3, and ensure that the signal strength and signal quality Q values of the upstream and downstream of the M90 window are greater than 60 (the larger the better). After confirmation, the downstream (DN) sensor is fixed with steel straps.
Standard Packing List:
(1) TUF-2000B host 1 piece
(2) TS-2 Sensor 1 pair
(3) Clamp Fixture 1 set
(4) Thermoplastic tube 1 set
(5) 5m Transducer Cable 1 pair
(6) English User Manual 1 piece
Notes:
You should apply the silicone gel to the contact surface of the sensor and pipe to prevent air interference. This is a very necessary step. Otherwise, the device will not receive a good signal.
Because silica gel is colloid, it can`t pass the security check of logistics company, so I can`t send it to you through logistics. I`m sorry! You can buy it at the local store. It`s very cheap.
You use silica gel similar to the following: